Temperature is often used among the information required to diagnose the state of a person’s internal organs, as pathological processes occurring in the human body are usually accompanied by thermal deviations not only on the surface but also inside the body. Modern medicine needs to improve non-invasive methods of measuring temperature deep in the human body for the purpose of early diagnosis and monitoring of various pathological processes. Medical radiothermography, in particular, allows non-invasive determination of the integral deep temperatures of internal organs.
The development of modern radiometric equipment involves solving such complex problems as ensuring high sensitivity of devices, matching the antenna to the biological object, ensuring the required penetration depth, and measuring absolute temperature with an error of at least 0.1°C. Given the importance of this problem, this paper demonstrates the approaches to the development of modern radiothermographs that the authors used in the process of creating a device for diagnosing breast temperature abnormalities.
The study considers the principles and schemes of construction of radiothermographic devices, requirements to the equipment as a whole, as well as to its main components. The authors conclude that it is optimal to use a modulation-compensation scheme of a radiometer for medical radiothermometry based on the measurement of two parameters: the integrated temperature and the integrated coefficient of reflection of electromagnetic waves from a biological object. A schematic diagram of a prototype radiothermograph, as well as its composition and design, is presented.
The approaches to the creation of modern radiothermographs presented in this work should be useful for developers of medical equipment for surgery, otorhinolaryngology, orthopedics, pediatrics, gynecology, etc., where measuring the distribution of patient body temperatures is of diagnostic value.

Keywords: microwave, radiometer, temperature, antenna, biological tissue, reflection, radiation.

Received 15.11 2023
References
  1. Esepkina N.A., Korol'kov D.V., Pariysky Yu.N. [Radio telescopes and radiometers] Radioteleskopy i radiometry. M., Nauka, 1973. (Rus)
  2. Troitsky V.S., Arrangereev E.A., Gustov A.B. et al. [On the possibility of using one’s own radio-thermal microwave radiation to measure the temperature of internal organs]. Uspekhi Fizicheskikh Nauk, 1981 vol. 134, iss. 1, pp. 155 - 158. (Rus)
  3. Shteinshleyger V.B., Misezhnikov G.S., Selsky A.G. [About one radiophysical method for detecting temperature anomalies in human internal organs]. Uspekhi Fizicheskikh Nauk, 1981 vol. 134, iss. 1, pp. 163 - 164. (Rus)
  4. Enander B., Larson G. Microwave radiometric measurements of the temperature inside a body. Electronics Letters, vol. 10, no. 15, pp. 317 - 317, 1974. https://doi.org/10.1049/el:19740250
  5. Asimakis N.P., Karanasiou I.S., Uzunoglu N.K. Non-invasive microwave radiometric system for intracranial applications: A study using the conformal l-notch microstrip patch antenna. Progress In Electromagnetics Research, 2011, vol. 117, pp. 83 - 101. https://doi.org/10.2528/PIER10122208
  6. Sugiura T., Hirata H., Hand J.W. et al. Five-band microwave radiometer system for noninvasive brain temperature measurement in newborn babies: Phantom experiment and confidence interval. Radio Science, 2011, vol. 46, iss. 5. https://doi.org/10.1029/2011RS004736
  7. Scheeler R., Kuester E.F., Popovic Z. Sensing depth of microwave radiation for internal body temperature measurement. IEEE Trans. Antennas Propag. 2014, vol. 62, iss. 3, pp. 1293 - 1303. https://doi.org/10.1109/TAP.2013.2295595
  8. Stauffer P.R., Rodriques D.B., Salahi S. et al. Stable microwave radiometry system for long term monitoring of deep tissue temperature. Proc. SPIE 8584, Energy-based Treatment of Tissue and Assessment VII, 85840R, 2013. https://doi.org/10.1117/12.2003976
  9. Hand J. W., Leeuwen G. M. J. V., Mizushina S. et al. Monitoring of deep brain temperature in infants using multi-frequency microwave radiometry and thermal modelling. Physics in Medicine and Biology, 2001, vol. 46, no. 7, pp. 1885 - 1903. https://doi.org/10.1088/0031-9155/46/7/311
  10. Maruyma K., Mizushina S., Sugiura T. et al. Feasibility of noninvasive measurement of deep brain temperature in newborn infants by multifrequency microwave radiometry. IEEE Transactions on Microwave Theory and Techniques, 2000, vol. 48, no. 11, pp. 2141 - 2147. https://doi.org/10.1109/22.884206
  11. Toutouzas K., Benetos G., Koutagiar I. et al. Noninvasive detection of increased carotid artery temperature in patients with coronary artery disease predicts major cardiovascular events at one year: Results from a prospective multicenter study. Atherosclerosis, 2017, vol. 262, 25 - 30. https://doi.org/10.1016/j.atherosclerosis.2017.04.019
  12. Macdonald A., Land D., Sturrock R. Microwave thermography as a noninvasive assessment of disease activity in inflammatory arthritis. Clinical Rheumatology, vol. 13, pp. 589 - 592, 1994. https://doi.org/10.1007/BF02242999
  13. Toutouzas K., Grassos C., Drakopoulou M. et al. First in vivo application of microwave radiometry in human carotids. Journal of the American College of Cardiology, 2012, vol. 59, no. 18, pp. 1645 - 1653. https://doi.org/10.1016/j.jacc.2012.01.033
  14. Pentazos G., Laskari K., Prekas K. et al. Microwave radiometry-derived thermal changes of small joints as additional potential biomarker in rheumatoid arthritis. JCR: Journal of Clinical Rheumatology, 2018, vol. 24, iss. 5, pp. 259 - 263. https://doi.org/10.1097/RHU.0000000000000719
  15. Toutouzas K., Synetos A., Nikolaou C. et al. Microwave radiometry: a new non-invasive method for the detection of vulnerable plaque. Cardiovascular Diagnosis and Therapy, vol. 2, no. 4, 2012. https://doi.org/10.3978/j.issn.2223-3652.2012.10.09
  16. Planck M. [Selected Works] Izbrannyye trudy. M., Nauka, 1975, 788 p. (Rus)
  17. Skripnik Yu.A., Yanenko A.F., Manoilov V.F. et al. [Microwave radiometry of physical and biological objects] Mikrovolnovaya radiometriya fizicheskaikh i biologicheskikh ob”yektov. Zhitomir, Publishing house “Volyn”, 2003, 408 p. (Rus)
  18. Troitsky V.S. On the theory of contact radiometric measurements of the internal temperature of bodies. Izvestija VUZov. Radiofizika, 1981, vol. 24, no. 9, p. 1054 - 1061. (Rus)
  19. Luedeke K.M., Sch³ñk Â., Koehler J. Radiation balance microwave thermograph for industrial and medical applications. Electronics Letters, 1978, vol. 14, iss. 6, ðp. 194 - 196. https://doi.org/10.1049/el:19780129
  20. Luedeke K.M., Koehler J. Microwave radiometric system for biomedical ‘true temperature’ and emissivity measurements. Microwave Power, 1983, vol. 18, iss. 3, ðp. 277 - 283. https://doi.org/10.1080/16070658.1983.11689332
  21. Gaevsky B.S., Marechek S.V., Meshkov Yu.V. et al. [Radiothermoscope]. Electronic Industry, 1987, no. 1, p. 29. (Rus)
  22. Kolisnichenko M.V., Kutsenko V.P., Skripnik Yu.A., Yanenko A.F. Microwave radiometry: medical aspects of use. Technology and design in electronic equipment, 2003, no. 5, pp. 23 - 25. (Rus)