CdZnTe-based gamma-ray detection module for radiation monitoring systems

  • V. A. Mokritsky Odessa National Polytechnic University, Ukraine
  • O. V. Maslov Odessa National Polytechnic University, Ukraine
  • Yu. E. Nikolaenko Odessa National Polytechnic University, Ukraine
  • M. V. Maksimov Odessa National Polytechnic University, Ukraine
Keywords: gamma-ray detection unit, CdZnTe detector, dependence of detector sensitivity on gamma-ray energy

Abstract

The use of CdZnTe for dose rate measurement is limited by the fact that the detector’s sensitivity varies by a factor of more than 10 in the γ-ray energy range of 10 keV to 3 MeV. To reduce the error caused by the detector sensitivity’s dependence on gamma-ray energy, digital correction of the output signal is used. The correction is performed by varying the pulse frequency at the output of the detection unit depending on the gamma-ray energy. A piecewise-linear linearization of the dependence of the count rate change coefficient on energy is applied.

Published
2005-06-30
How to Cite
Mokritsky, V. A., Maslov, O. V., Nikolaenko, Y. E., & Maksimov, M. V. (2005). CdZnTe-based gamma-ray detection module for radiation monitoring systems. Technology and Design in Electronic Equipment, (3), 15-17. Retrieved from https://www.tkea.com.ua/index.php/journal/article/view/TKEA2005.3.15