Calculation of characteristics of X-ray devices

  • A. N. Orobinskyi National Scientific Centre "Institute of Metrology", Kharkiv, Ukraine
Keywords: X-ray device, X-ray tube, X-ray radiation, anode voltage, anode current, spectral resolution, mean photon energy, air kerma power

Abstract

Actuality of this work is related to human radiation safety during tuning and regulation of X-ray devices in the process of their development and production. The more precise the calculations for the device are, the less time is required for its tuning and regulation, and thus people are less exposed to radiation. When developing an X-ray device, it is necessary to choose an X-ray tube and filters taking into account the application domain of the device. In order to do this, one should know anode voltage, X-ray tube anode current, material and thickness of filters, i.e. to calculate these characteristics at the set quality of X-ray radiation. The known published studies do not give any solution to this problem.
The scientific novelty of this work is that it establishes the interdependence between main characteristics of the X-ray device: the function of the device defines the quality of X-ray radiation (mean photon energy and air kerma power); mean photon energy depends on the X-ray anode tube voltage and spectral resolution; air kerma power depends on anode tube voltage, current of X-ray tube anode, spectral resolution, thicknesses of the filters and their materials; spectral resolution depends on thicknesses of filters and their materials; thickness of filters depends on the material of the filter (the linear coefficient of weakening of X-ray radiation). Knowledge of interdependence of basic characteristics of the X-ray devices allowes developing simple algorithm for their calculation at the values of homogeneity coefficient from 0,8 to 1, which makes it possible to choose an X-ray tube and filters with the purpose of obtaining X-ray radiation of the set quality.

References

Dushkin S.A., Ivanskyi V.B., Kurov A.M., Odinets V.A., Orobinskyi A.N. The measurement errors of X-ray devices features. Tekhnologiya i konstruirovanie v elektronnoi apparature, 2011, no. 3, pp. 44-46. (Rus)

IEC 61267:2005-11. Medical diagnostic X-ray equipment. Radiation conditions for use in the determination of characteristics.

DIN 6809-4:1987-04. Clinical dosimetry. Applications of X-rays with peak voltages between 10 and 100 kV in radiotherapy and soft tissue diagnostics.

DIN 6809-5:1996-02. Clinical dosimetry. Part 5: Application of X-rays with peak voltages between 100 and 400 kV in radiotherapy.

Dushkin S.A., Kurov A.M., Odinets V.A., Orobinskyi A.N. The calculation of features of X-ray radiation. Tekhnologiya i konstruirovanie v elektronnoi apparature, 2008, no. 6, pp. 60-64. (Rus)

Dushkin S.A., Kurov A.M., Odinets V.A., Orobinskyi A.N. [Researches of metrology features of X-ray device type of PANTAK HF160]. Proc. of the 6th Int. sc.-tech. conf. “Metrology and measuring technique”, vol. II, Kharkov, 2008, pp. 261-265. (Rus)

Dushkin S.A., Kurov A.M., Odinets V.A., Orobinskyi A.N. [Radiation output of X-ray device]. Proc. of the 7th Int. sc.-tech. conf. “Metrology and measuring technique”, vol. II, Kharkov, 2010, pp. 253-257. (Rus)

DSTU ISO 4037-1:2006. Standartne rentgenivs’ke ta gamma-vyprominennya dlya kalibruvannya dozymetriv i vymiryuvachiv potuzhnosti dozy ta dlya vyznachennya iikh chutlyvosti v zalezhnosti vid energiii fotoniv. Ch. 1. Kharakterystyky vyprominen’ ta metody, shcho iikh stvoryuyut’ [State Standard ISO 4037-1:2006, part 1]. (Rus)

[X-rays: Reference book]. Ed. by Ì.À. Blohin, Moskow, Izdatel’stvo inostrannoi literatury, 1960. (Rus)

Nemec O.F., Gofman Y.V. Reference book on nuclear physics. Kiev, Naukova dumka, 1975. (Rus)

Published
2015-12-25
How to Cite
Orobinskyi, A. N. (2015). Calculation of characteristics of X-ray devices. Technology and Design in Electronic Equipment, (5–6), 55-63. https://doi.org/10.15222/TKEA2015.5-6.55